$33
Oscilloscope mode Frequency response – dc coupled Without probes and test leads (with BB120
Oscilloscope mode
Frequency response – dc coupled
Without probes and test leads (with BB120)
123B: dc to 20 MHz (-3 dB)
124B and 125B: dc to 40 MHz (-3 dB)
DC to 12.5 MHz (-3 dB) / dc to 20 MHz (-6 dB)
123B: dc to 20MHz (-3 dB)
124B and 125B: dc to 40 MHz (-3 dB)
Frequency response – ac coupled (lf roll off)
Without probes and test leads
<10 Hz (-3 dB)
<10 Hz (-3 dB)
<10 Hz (-3 dB)
Rise time, excluding probes,
test leads
123B <17.5 ns
124B and 125B <8.75 ns
Input impedance
Without probes and test leads
1 MΩ//20 pF
1 MΩ//24 pF
1 MΩ//230 pF
5 MΩ//15.5 pF
Sensitivity
5 mV to 200 V/div
Analog bandwidth limiter
10 kHz
Display modes
A, -A, B, -B
Max. input voltage A and B
Direct, with test leads, or with VP41 Probe
600 Vrms Cat IV, 750 Vrms maximum voltage.
600 Vrms
Max. floating voltage, from any terminal to ground
600 Vrms Cat IV, 750 Vrms up to 400Hz
Scope modes
Normal, Single, Roll
Ranges (normal)
Equivalent sampling
123B: 20 ns to 500 ns/div,
1 μs to 5 s/div
1 μs to 5 s/div
1s to 60 s/div
Sampling rate (for both channels simultaneously)
Equivalent sampling (repetitive signals)
Up to 4 GS/s
40 MS/s
Trigger
Screen update
Free run, on trigger
Source
A, B
Sensitivity A and B
@ DC to 5 MHz
0.5 divisions or 5 mV
123B: 4 divisions
123B: N/A
Slope
Positive, negative
Advanced scope functions
Display modes
Normal
Captures up to 25 ns glitches and displays analog-like persistence waveform
Suppresses noise from a waveform
Does not capture glitches between samples
Records and displays the minimum and maximum of waveforms over time
Auto set (Connect-and-View™)
Continuous fully automatic adjustments of amplitude, time base, trigger levels, trigger gap, and hold-off. Manual override by user adjustment of amplitude, time base, or trigger level.
Dual input meter
Input A and input B
Ranges
500 mV, 5 V, 50 V, 500 V, 750 V
Accuracy
±(0.5% +5 counts)
Common mode rejection (CMRR)
>100 dB @ dc, >60 dB @ 50, 60, or 400 Hz
Full scale reading
5000 counts
Ranges
500 mV, 5 V, 50 V, 500 V, 750 V
Accuracy for 5% to 100% of range (DC coupled)
DC to 60 Hz (V ac+dc)
±(1% +10 counts)
±(1% +10 counts)
Accuracy for 5% to 100% of range (AC or dc coupled)
60 Hz to 20 kHz
±(2.5% +15 counts)
DC rejection (only VAC)
>50 dB
Common mode rejection (CMRR)
>100 dB @ dc
Full scale reading
5000 counts, reading is independent of any signal crest factor.
Modes
Max peak, Min peak, or pk-to-pk
Ranges
500 mV, 5 V, 50 V, 500 V, 2200 V
Accuracy
Accuracy Max peak or Min peak
5% of full scale
10% of full scale
Full scale reading
500 counts
Ranges
123B: 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,1 MHz, 10 MHz, and 50 MHz
Frequency range
15 Hz (1 Hz) to 50 MHz in continuous autoset
Accuracy @1 Hz to 1 MHz
±(0.5% +2 counts)
Full scale reading
10,000 counts
Max reading
50.00 kRPM
Accuracy
±(0.5% +2 counts)
Range
2% to 98%
Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset
Frequency range
15 Hz (1 Hz) to 30 MHz in continuous autoset
Full scale reading
1000 counts
With current clamp
Ranges
Same as V dc, V ac, V ac+dc, or PEAK
0.1 mV/A, 1 mV/A, 10 mV/A, 100 mV/A, 400 mV/A, 1 V/A, 10 mV/mA
Same as V dc, V ac, V ac+dc, or PEAK (add current clamp uncertainty)
Range
200 °C/div (200 °F/div)
Scale factor
1 mV/°C and 1 mV/°F
Accuracy
As V dc (add temp. probe uncertainty)
0 dBV
1 V
0 dBm (600 Ω /50 Ω)
1 mW referenced to 600 Ω or 50 Ω
dB on
V dc, V ac, or Vac+dc
Full scale reading
1000 counts
Range
1 to 10
Full scale reading
90 Counts
Modes
A to B, B to A
Range
0 to 359 degrees
Resolution
1 degree
Configurations
1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET mode only)
Power factor (PF)
Ratio between watts and VA range – 0.00 to 1.00
Watt
RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes)
999 counts
VA
Vrms x Arms
999 counts
VA reactive (var)
√((VA)2-W2)
999 counts
Purpose
To measure on pulse width modulated signals, like motor drive inverter outputs
Principle
Readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency
Accuracy
As Vrms for sinewave signals
Input A to common
Ranges
123B and 124B
500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
50 Ω, 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
Accuracy
±(0.6% + 5 counts) 50 Ω ±(2% + 20 counts)
Full scale reading
50 Ω to 5 MΩ – 5000 counts, 30 MΩ – 3000 counts
Measurement current
0.5 mA to 50 nA, decreases with increasing ranges
Open circuit voltage
<4 V
Beep
<(30 Ω ±5 Ω) in 50 Ω range
Measurement current
0.5 mA
Detection of shorts of
≥1 ms
Measurement voltage
@0.5 mA
>2.8 V
<4 V
Measurement current
0.5 mA
Polarity
+ on input A, – on COM
Ranges
50 nF, 500 nF, 5 μF, 50 μF, 500 μF
Full scale reading
5000 counts
Measurement current
500 nA to 0.5 mA, increases with increasing ranges
Advanced meter functions
Zero Set
Set actual value to reference
AutoHold (on A)
Captures and freezes a stable measurement result. Beeps when stable. AutoHold works on the main meter reading, with thresholds of 1 Vpp for AC signals and 100 mV for DC signals.
Fixed decimal point
Activated by using attenuation keys
Cursor Readout (124B and 125B)
Sources
A, B
Single vertical line
Dual vertical lines
Peak-peak, time distance and reciprocal time distance readout
Dual horizontal lines
High, low and peak-peak readout
Rise or fall time
Transition time, 0%-level and 100%-level readout (manual or auto leveling; auto leveling only
possible in single channel mode)
Accuracy
As oscilloscope accuracy
Recorder
Measurement Speed
Maximum 2 measurement/s
Record Size (min, max, average)
2 M readings for 1 channel
Recorded Time Span
2 weeks
Maximum number of events
1024
Maximum sample rate
400 K sample/s
Size Internal memory
400 M samples Recorded Time
Span internal memory
15 minutes at 500 μs/div
11 hours at 20 ms/div
Record Size SD card
1.5 G samples
Recorded Time Span SD card
11 hours at 500 μs/div
14 days at 20 ms/div
Maximum number of events
64
Power Quality (125B only)
Readings
Watt, VA, var, PF, DPF, Hz
Watt, VA, var ranges (auto)
250 W to 250 MW, 625 MW, 1.56 GW
±(2% + 6 counts)
±(4% + 4 counts)
DPF
0.00 to 1.00
PF
0.00 to 1.00, ±0.04
Frequency range
10.0 Hz to 15.0 kHz
40.0 Hz to 70.0 Hz
Number of Harmonics
DC to 51
Readings / Cursor readings
(fundamental 40 Hz to 70 Hz)
Vrms / Arms /Watt
Each harmonic from fundamental maybe selected for individual readings
BusHealth tester (Fluke 125B only)
Type
Subtype
Protocol
AS-i
NEN-EN50295
CAN
ISO-11898
Interbus S
RS-422
EIA-422
Modbus
RS-232
RS-232/EIA-232
RS-485/EIA-485
Foundation Fieldbus
H1
61158 type 1, 31.25 kBit
Profibus
DP
EIA-485
61158 type 1
Miscellaneous
Display
Type
5.7-inch color active matrix TFT
640 x 480 pixels
Waveform Display
Vertical
10 div of 40 pixels
12 div of 40 pixels
Power
External
Via Power Adapter BC430
10 V DC to 21 V DC
5 W typical
5 mm jack
Via Battery Pack BP290
Rechargeable Li-Ion 10.8 V
7 hours with 50% backlight brightness
4 hours with test tool off, 7 hours with test tool on
0 to 40 °C (32 to 104 °F) during charging
Memory
Internal memory can store 20 data sets (screen waveform and setup)
Micro SD card slot with optional SD card (max size of 32 GB)
Mechanical
Size
259 x 132 x 55 mm
(10.2 x 5.2 x 2.15 in)
1.4 kg (3.2 lb) including battery pack
Interface
Optically isolated
Transfer screen copies (bitmaps), settings and data
OC4USB optically isolated USB adapter/cable, (optional), using FlukeView® software for Windows®.
Fast transfer of screen copies (bitmaps), settings and data to PC/laptop, tablet, smartphone, etc. A USB port is provided for attaching the WiFi dongle. Do not use the USB port with a cable for safety reasons.
Environmental
Environmental
MIL-PRF-28800F, Class 2
Temperature
Battery Operation
0 to 40 °C (32 to 104 °F)
0 to 50 °C (32 to 122 °F)
-20 to 60 °C (-4 to 140 °F)
Humidity (Operating)
@ 0 to 10 °C (32 to 50 °F)
Non-condensing
95%
75%
45%
Storage
@ -20 to 60 °C (-4 to 140 °F)
Non-condensing
Altitude
Operating at 3 km (10,000 feet)
CAT III 600 V
CAT IV 600 V
12 km (40,000 feet)
EMC electromagnetic compatibility
International
IEC 61326-1: Industrial, CISPR 11: Group 1, Class A
Class A Equipment (Industrial Broadcasting & Communication Equipment)
47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
Wireless radio with adapter
Frequency range
2412 MHz to 2462 MHz
<100 mW
Enclosure protection
IP51, ref: EN/IEC60529
Safety
General
IEC 61010-1: Pollution Degree 2
IEC 61010-2-033: CAT IV 600 V/CAT III 750 V
Max. input voltage input A and B
Direct on input or with leads
600 Vrms CAT IV for derating
600 Vrms for derating
600 Vrms Cat IV, 750 Vrms up to 400 Hz